Baumer Standard RTD temperature sensor

Housing DIN form B Immersion depth to 3000 mm 4 … 20 mA or HART output Pt100 class A/B, Pt1000-configurable

Interested with this product

Technical data
Performance characteristics
Process temperature
  • -50 … 400 °C
  • -50 … 600 °C , optional
Max. flow velocity
  • 40 m/s , gases
  • 5 m/s , liquids
Pt100 accuracy class (EN 60751)
  • B (± 0.3 °C at 0 °C)
  • ± (0.3 + 0.005 × t)°C
  • A (± 0.15 °C at 0 °C)
  • ± (0.15 + 0.002 × t)°C
  • 1/3 B (± 0.1 °C at 0 °C)
  • ± 1/3 × (0.3 + 0.005 × t)°C
  • 1/6 B (± 0.05 °C at 0 °C)
  • ± 1/6 × (0.3 + 0.005 × t)°C
Pt1000 accuracy class (EN 60751)
  • B (± 0.3 °C at 0 °C)
  • ± (0.3 + 0.005 × t)°C
  • 1/3 B (± 0.1 °C at 0 °C)
  • ± 1/3 × (0.3 + 0.005 × t)°C
Response time, T50
  • ≤ 1.5 s , Ø4 mm
  • ≤ 6.1 s , Ø6 mm
  • ≤ 7.6 s , Ø8 mm
  • ≤ 11.1 s , Ø10 mm
Process connection
Sensor length
  • 20 … 3000 mm
Sensor tube material
  • AISI 316L (1.4404)
Surface roughness wetted parts
  • Ra ≤ 0.8 µm
Ambient conditions
Degree of protection (EN 60529)
  • IP 65
Humidity
  • ≤ 100 % RH , condensing
Vibration (sinusoidal) (EN 60068-2-6)
  • 1.6 mm p-p (2 … 25 Hz), 4 g (25 … 100 Hz), 1 octave / min.
Output signal
Without transmitter
  • 1 x Pt100, 2-wire
  • 1 x Pt100, 4-wire
  • 2 x Pt100, 2-wire
  • 1 x Pt1000, 2-wire
With transmitter
  • 4 … 20 mA , 2-wire
  • 4 … 20 mA , + HART®
Compliance and approvals
Railway applications
  • EN 50155
Explosion protection
  • ATEX II 1G Ex ia IIC T4/T5
  • ATEX II 3G Ex nA IIC T5
  • Ex ia Simple apparatus, gas and dust
Installation and operating instructions
Certificates / approvals
Technical Reports / White Paper
Software
3D CAD

Equivalent Products